(for use as a puncture or injection needle), 1 of Each
Tray Contents:
1 CSR Outer Wrap 28 x 28
1 Prep Tray 10x5x2
1 MRI "Non-Ferromagnetic" Disposable Titanium #11 Scalpel - This scalpel is safe to use for any interventional procedure for the initial puncture of the skin
1 MRI "Non-Ferromagnetic" Needle 20g 9cm - This needle is completely safe for use as a puncture or injection needle. Non-Clinical testing demonstrated that this product is MR conditional and can be used in the MRI environment according to the following conditions:
1. Static magnetic of 3-Tesla or less
2. Highest spatial gradient magnetic field of 720-Gauss/cm or less
Important note: These products are intended for use inside of the MRI environment (in the MR system room). They should not be utilized directly inside of the MR system (inside the bore of the scanner) during its operation (scanning). As such, the assessment of magnetic field interactions for this product specifically involved evaluations of translational attraction and torque in relation to exposure to a 3-Tesla MR system. This product has not been tested for MRI-related heating and artifacts.